A Reinforced Endotracheal Tube Completely Severed by a Patient Bite and Lodged in the Right Main Bronchus / 대한구급학회지
The Korean Journal of Critical Care Medicine ; : 70-73, 2017.
Article in English | WPRIM | ID: wpr-770975
ABSTRACT
Reinforced endotracheal tubes (ETTs) are designed to resist kinking or compression. However, these have a potential risk of being obstructed or severed by a patient's bite. We report a case in which a reinforced ETT was severed by tube-bite while the patient was in the prone position during an intensive care unit stay. Bronchoscopic evaluation showed that the severed distal part of the tube had lodged in the patient's right main bronchus, and it had to be surgically removed. The patency of reinforced ETTs should be carefully monitored in patients intubated in the prone position.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Bronchi / Bronchoscopy / Prone Position / Patient Rights / Airway Obstruction / Intensive Care Units / Intubation Limits: Humans Language: English Journal: The Korean Journal of Critical Care Medicine Year: 2017 Type: Article