Optimization of Soft Endoscope Cleaning and Disinfection Equipment / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 384-387, 2019.
Article
in Chinese
| WPRIM
| ID: wpr-772478
ABSTRACT
According to the national soft endoscope cleaning standard and the difference between the previous and latest version, the origin decontamination equipment is upgraded. Sensory control monitoring results in the last year shows that the average cleaning time for a single endoscope has been reduced from 20 min 55 s to 17 min 16 s. The number of bacterial colony has been reduced from 6.8 to 3.1. The pass rate of sensory control monitoring keeps steady. The upgraded decontamination equipment not only meets the requirements of the latest version of the standard,but also improves the efficiency and effectiveness of electronic endoscope disinfection.