Research on Fault Risk Identification and Control of CT Simulator Based on FMEA / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
; (6): 303-306, 2019.
Article
in Zh
| WPRIM
| ID: wpr-772500
Responsible library:
WPRO
ABSTRACT
OBJECTIVE@#To explore the high-risk fault risk of CT simulator and the main causes of the risk, and to put forward effective risk management strategies.@*METHODS@#The failure mode and effect analysis method was used to identify and control the operational fault risk of CT simulator.@*RESULTS@#5 major fault components, 8 fault failure models and 17 failure causes were analyzed. The top 5 failure causes are:anode target surface burn caused by direct scanning without warming up the tube (590.4), tube failure (518.2), burnout of joints caused by aging of high voltage cables (424.2), motor carbon brush wear (304.8) and belt break (296.4).@*CONCLUSIONS@#The failure mode and effect analysis method can effectively identify the risk of equipment failure, and thus specifically formulate risk management and control measures to ensure the normal operation of equipment and the safety of doctors and patients.
Key words
Full text:
1
Index:
WPRIM
Main subject:
Reference Standards
/
Risk Management
/
Tomography, X-Ray Computed
/
Equipment Failure
/
Methods
Type of study:
Diagnostic_studies
/
Etiology_studies
Limits:
Humans
Language:
Zh
Journal:
Chinese Journal of Medical Instrumentation
Year:
2019
Type:
Article