Your browser doesn't support javascript.
loading
Development of the Extraction Reference Point Beam Diagnostic System for Proton Medical Accelerator / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 102-105, 2019.
Article in Chinese | WPRIM | ID: wpr-772554
ABSTRACT
er to detect the beam quality of the SC200 superconducting cyclotron,measure the beam at the extraction reference and the acceptance of the accelerator is realized.This article mainly introduces the design that use the scintillation screen at the extraction reference to measure the beam profile,position and use the Faraday cup to measure the current intensity with 2.5 level accuracy.The remoted controlling of probes and the acquisition and processing of signal based on LabVIEW and PLC.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Proton Therapy Type of study: Diagnostic study Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2019 Type: Article

Similar

MEDLINE

...
LILACS

LIS