Your browser doesn't support javascript.
loading
Comparative Cardiac Phantom Study Using Tc-99m/I-123 and Tl-201/I-123 Tracers with Cadmium-Zinc-Telluride Detector-Based Single-Photon Emission Computed Tomography / 대한핵의학회잡지
Korean Journal of Nuclear Medicine ; : 57-63, 2019.
Article in English | WPRIM | ID: wpr-786446
ABSTRACT

OBJECTIVE:

A recently introduced single-photon emission computed tomography (SPECT), based on cadmium-zinc-telluride (CZT) detectors (D-SPECT), supports high energy resolution for cardiac imaging. Importantly, the high energy resolution may allow simultaneous dual-isotope (SDI) imaging (e.g., using Tc-99m and I-123). We quantitatively evaluated Tc-99m/I-123 SDI imaging by D-SPECT in comparison with conventional T1-201/I-123.MATERIALS AND

METHODS:

Energy resolution was measured as a percentage of the full width at half maximum(FWHM) for Tc-99m, I-123, and Tl-201. The impact of cross-talk and reconstructed image contrast were quantified by measuring the contrast-to-noise ratio (CNR), and the transmural defect contrast in the left ventricle wall (CTD) induced by a difference in energy, for combinations of Tc-99m/I-123 or Tl-201/I-123, using an RH-2 cardiac phantom. Corresponding measurement was also carried out in Anger SPECT (A-SPECT).

RESULTS:

The energy resolution of the D-SPECTsystem was 5.4%/5.1%for Tc-99m/I-123 and 5.4%/5.3%for Tl-201/I-123, which was approximately two times higher than the A-SPECT. No notable difference was confirmed in the CNRs of the two systems, but T1-201/I-123 showed overall higher value than Tc-99m/I-123. Compared to A-SPECT, CTD of D-SPECT significantly increased with both Tc-99m/I-123 and T1-201/I-123 (p < 0.05). In DSPECT, the combination of Tc-99m/I-123 had a slightly better CTD than T1-201/I-123. In addition, CTD of Tc-99m/I-123 was improved with scatter correction at both nuclides (p < 0.05), but in Tl-201/I-123, no significant improvement was confirmed in I-123 (p > 0.05).

CONCLUSION:

D-SPECT was considered to be capable of performing high-quality SDI imaging using Tc-99m/I-123.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Tomography, Emission-Computed / Tomography, Emission-Computed, Single-Photon / Heart Ventricles / Anger Language: English Journal: Korean Journal of Nuclear Medicine Year: 2019 Type: Article