Your browser doesn't support javascript.
loading
Applications of the Fe3O4 nanocomposite modified by Humic Acid in determination of S-phenylmercapturic acid in urine by dispersive solid-phase extraction and HPLC / 中华劳动卫生职业病杂志
Chinese Journal of Industrial Hygiene and Occupational Diseases ; (12): 463-465, 2018.
Article in Chinese | WPRIM | ID: wpr-806622
ABSTRACT
Objective@#To establish a method for determination the S-phenylmercapturic acid in urine by dispersive solid-phase extraction using Humic Acid/Fe3O4 magnetic nanocomposite as adsorbent.@*Methods@#The 5 ml of urine samples were adjusted to pH 1.0 and extracted by Fe3O4@HA. Then the analytes were separated on EC-C18 capillary column and detected by HPLC-VWD. The S-phenylmercapturic acid was characterized by the retention time and quantified by peak area and external standard method.@*Results@#The standard curves of SPMA showed significant linearity between 0.04~1.00 mg/L (r=0.999 7) . The average recovery was 94.2%~102.4%. The intra-day and inter-day precisions (RSD) were 2.9~6.7% (n=6) and 3.1~7.5% (n=6) respectively. The detect limit of SPMA was 0.012 g/L (S/N=3) .@*Conclusion@#This method is simple, rapid, sensitive and accurate. It is applicable for determination of SPMA in the urine of works who were exposed to benzene.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Industrial Hygiene and Occupational Diseases Year: 2018 Type: Article