Direct determination of manganese in urine by Graphite Furnace Atomic Absorption Spectrometry / 中华劳动卫生职业病杂志
Chinese Journal of Industrial Hygiene and Occupational Diseases
;
(12): 535-537, 2018.
Article
in Chinese
| WPRIM
| ID: wpr-806804
ABSTRACT
Objective@#To establish a method for the determination of manganese in urine by graphite furnace atomic absorption spectrometry (AAS) without the use of matrix modifier.@*Methods@#The urine samples were 5 times diluted with 1% nitric acid then directly determined by AAS. Zeeman was used for background correction.@*Results@#The linear range for determination of manganese in urine was 5~60 μg/L (urine) . The correlation coefficient was greater than 0.995 with the detection limit of 1.5 μg/L and with the lower limit of quantification of 5.0 μg/L. The relative standard deviations (RSDs) of within-run precision was between 1.1%~4.3%, the RSDs of between-run precision was between 3.3%~7.0%. The average recovery was 102.6%. The samples can be stored for 14 days at room temperature, 4℃, -8 ℃ and -35 ℃.@*Conclusion@#The method is feasible for determination of manganese in urine.