Your browser doesn't support javascript.
loading
Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study
Journal of Periodontal & Implant Science ; : 56-66, 2020.
Article in English | WPRIM | ID: wpr-811257
ABSTRACT

PURPOSE:

A stability-measuring device that utilizes damping capacity analysis (DCA) has recently been introduced in the field of dental implantology. This study aimed to evaluate the sensitivity and reliability of this device by measuring the implant stability of ex vivo samples in comparison with a resonance frequency analysis (RFA) device.

METHODS:

Six implant beds were prepared in porcine ribs using 3 different drilling protocols to simulate various implant stability conditions. Thirty-six pork ribs and 216 bone-level implants measuring 10 mm in height were used. The implant beds were prepared using 1 of the following 3 drilling protocols 10-mm drilling depth with a 3.5-mm-diameter twist drill, 5-mm drilling depth with a 4.0-mm-diameter twist drill, and 10-mm drilling depth with a 4.0-mm-diameter twist drill. The first 108 implants were external-connection implants 4.0 mm in diameter, while the other 108 implants were internal-connection implants 4.3 mm in diameter. The peak insertion torque (PIT) during implant placement, the stability values obtained with DCA and RFA devices after implant placement, and the peak removal torque (PRT) during implant removal were measured.

RESULTS:

The intraclass correlation coefficients (ICCs) of the implant stability quotient (ISQ) results obtained using the RFA device at the medial, distal, ventral, and dorsal points were 0.997, 0.994, 0.994, and 0.998, respectively. The ICCs of the implant stability test (IST) results obtained using the DCA device at the corresponding locations were 0.972, 0.975, 0.974, and 0.976, respectively. Logarithmic relationships between PIT and IST, PIT and ISQ, PRT and IST, and PRT and ISQ were observed. The mean absolute difference between the ISQ and IST values on a Bland-Altman plot was −6.76 (−25.05 to 11.53, P<0.05).

CONCLUSIONS:

Within the limits of ex vivo studies, measurements made using the RFA and DCA devices were found to be correlated under a variety of stability conditions.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Ribs / In Vitro Techniques / Dental Implants / Osseointegration / Torque / Diagnostic Techniques and Procedures / Red Meat Type of study: Diagnostic study Language: English Journal: Journal of Periodontal & Implant Science Year: 2020 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Ribs / In Vitro Techniques / Dental Implants / Osseointegration / Torque / Diagnostic Techniques and Procedures / Red Meat Type of study: Diagnostic study Language: English Journal: Journal of Periodontal & Implant Science Year: 2020 Type: Article