Your browser doesn't support javascript.
loading
Improvement of the Dose Calculation Accuracy Using MVCBCT Image Processing / 의학물리
Korean Journal of Medical Physics ; : 62-69, 2012.
Article in Korean | WPRIM | ID: wpr-81821
ABSTRACT
The dose re-calculation process using Megavoltage cone-beam CT images is inevitable process to perform the Adaptive Radiation Therapy (ART). The purpose of this study is to improve dose re-calculation accuracy using MVCBCT images by applying intensity calibration method and three dimensional rigid body transform and filtering process. The three dimensional rigid body transform and Gaussian smoothing filtering process to MVCBCT Rando phantom images was applied to reduce image orientation error and the noise of the MVCBCT images. Then, to obtain the predefined modification level for intensity calibration, the cheese phantom images from kilo-voltage CT (kV CT), MVCBCT was acquired. From these cheese phantom images, the calibration table for MVCBCT images was defined from the relationship between Hounsfield Units (HUs) of kV CT and MVCBCT images at the same electron density plugs. The intensity of MVCBCT images from Rando phantom was calibrated using the predefined modification level as discussed above to have the intensity of the kV CT images to make the two images have the same intensity range as if they were obtained from the same modality. Finally, the dose calculation using kV CT, MVCBCT with/without intensity calibration was applied using radiation treatment planning system. As a result, the percentage difference of dose distributions between dose calculation based on kVCT and MVCBCT with intensity calibration was reduced comparing to the percentage difference of dose distribution between dose calculation based on kVCT and MVCBCT without intensity calibration. For head and neck, lung images, the percentage difference between kV CT and non-calibrated MVCBCT images was 1.08%, 2.44%, respectively. In summary, our method has quantitatively improved the accuracy of dose calculation and could be a useful solution to enhance the dose calculation accuracy using MVCBCT images.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Orientation / Calibration / Cheese / Electrons / Cone-Beam Computed Tomography / Head / Lung / Neck / Noise Language: Korean Journal: Korean Journal of Medical Physics Year: 2012 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Orientation / Calibration / Cheese / Electrons / Cone-Beam Computed Tomography / Head / Lung / Neck / Noise Language: Korean Journal: Korean Journal of Medical Physics Year: 2012 Type: Article