Your browser doesn't support javascript.
loading
Evaluation of the Wisecheck Glucose Monitoring System
Laboratory Medicine Online ; : 15-21, 2014.
Article in Korean | WPRIM | ID: wpr-82417
ABSTRACT

BACKGROUND:

Glucometers are widely used for self-monitoring and point-of-care testing in diabetes management. We evaluated the performance of the recently developed Wisecheck Glucose Monitoring System (Wisemeditech, Korea) compared to that of 2 other well-known glucometer systems.

METHODS:

The Wisecheck glucometer was evaluated for precision, linearity, and carryover rate. One-hundred fifty samples samples were tested, and the results obtained from the Wisecheck glucometer, ACCU-CHEK Performa (Roche Diagnostics, Germany) and SD GlucoLink (SD Diagnostics, Korea) were compared to those obtained using the laboratory reference method from the Toshiba 200FR (Toshiba, Japan), according to the Clinical and Laboratory Standards Institute (CLSI) guidelines.

RESULTS:

The coefficient of variation (CV) values for within-run imprecision at low, middle, and high levels were 2.06%, 1.02%, and 2.02%, respectively, and the CV values for total-run imprecision at low, middle, and high levels were 2.98%, 2.41%, and 1.88%, respectively. In the linearity test, the coefficient of determination (R2) was 0.9985 in glucose concentration ranging from 48.6 mg/dL to 428 mg/dL (P<0.0001). The results obtained using the Wisecheck glucometer were well correlated with those obtained using the Toshiba 200FR (R2=0.980, P<0.0001). The carryover rate was 0.12%.

CONCLUSIONS:

The Wisecheck glucometer showed good precision, linearity, and correlation with the reference method. It provided rapid and reliable measurements of blood glucose levels and seemed appropriate for use in diabetes management.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Blood Glucose / Diabetes Mellitus / Glucose / Methods Type of study: Practice guideline Language: Korean Journal: Laboratory Medicine Online Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Blood Glucose / Diabetes Mellitus / Glucose / Methods Type of study: Practice guideline Language: Korean Journal: Laboratory Medicine Online Year: 2014 Type: Article