Your browser doesn't support javascript.
loading
Versa HD Volume-modulated Accelerator High Voltage Circuit Fault Maintenance / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 371-373, 2020.
Article in Chinese | WPRIM | ID: wpr-828184
ABSTRACT
This paper introduces a fast repair methods of Versa HD volume-modulated accelerator's high voltage circuit faultthe key points test method. To identify five key points①Enter maintenance mode to check for AFC deviation; ② The magnetron waveform MI and PFN waveform PFN V are detected on the maintenance terminal board; ③ Detect thyratron waveform; ④ Check the supply voltage of thyratron; ⑤ HT PSU 600 V DC test 600 V normal or not. The waveform or voltage is also measured to efficiently narrow the fault range to find out the cause of the fault, quickly remove the fault.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Maintenance Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2020 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Maintenance Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2020 Type: Article