ABSTRACT
Objective:
To establish a flow-injection chemiluminescence method for the determination of doxorubicin, epirubicin and mitoxantrone and study its reaction mechanism.Methods:
In alkaline medium, chemiluminescence of luminol-potassium permanganate system could be inhibited obviously by anthracycline antibiotics. Combined with flow-injection technique, a new chemiluminescence method for determining the anthracycline antibiotics was set up. The chemiluminescence mechanism of the luminol-potassium permanganate system was also discussed.Results:
Under optimal conditions, the good linear ranges of doxorubicin, epirubicin and mitoxantrone were 5.0×10-9-1.0×10 -6 g/mL, 1.0×10-9-1.0×10-5 g/mL and 3×10-9-1.0×10-6 g/mL, respectively. The detection limits of doxorubicin, epirubicin and mitoxantrone were 3.0×10-9 g/mL, 5.0×10-6 g/mL and 2.0×10-9 g/mL, respectively. During eleven repeated inter-day and intra-day precision tests of 1.0×10-6 g/mL samples, the relative standard deviations corresponded to reference values of 3.0% ,2.8% and 2.1%.Conclusion:
The developed method is sensitive, accurate, rapid and of low cost. It can be applied to determine doxorubicin hydrochloride, epirubicin hydrochloride and mitoxantrone hydrochloride in injection preparations.