Impact of MRI fringe magnetic field on radiation beam characteristics of electron guns in real time MRI-guided radiotherapy / 中国医学影像技术
Chinese Journal of Medical Imaging Technology ; (12): 433-438, 2019.
Article in Chinese | WPRIM | ID: wpr-861441
ABSTRACT

Objective:

To investigate the impact of MR fringe magnetic field on the electron beam characteristics of the electron gun in the MRI-guided radiotherapy (MRIgRT). Methods Electromagnetic field analysis software was used to simulate the beam characteristics of both Litton L2087 electron gun and Varian 600C electron gun under different magnetic field strengths for the in-line MRIgRT configuration. Results Two electron guns exhibited qualitatively similar behavior in the fringe magnetic field. As field strength increased, the current emitted by the cathode increased slowly, which could increase up to 2.72% and 1.93% from the original value without external magnetic field for Litton L2087 and Varian 600C electron gun, respectively. The current at the anode exit started to decrease after the initial plateau, and after reaching a minimum value, the current started to increase slowly, and the current loss could reach up to 80.25% and 73.48% of the initial value, respectively. At high values of the field strength, the beam radius changed periodically. Beam root mean square emittance fluctuated under different magnetic field strengths. Conclusion For in-line MRIgRT configuration, the fringe magnetic field can affect the trajectory of the beam and cause a large amount of current loss in the electron gun, and the beam RMS emittance varies greatly.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Medical Imaging Technology Year: 2019 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Medical Imaging Technology Year: 2019 Type: Article