Rapid determination of iodine in whole blood by direct alkali dilution ICP-MS / 中华地方病学杂志
Chinese Journal of Endemiology ; (12): 606-609, 2020.
Article in Chinese | WPRIM | ID: wpr-866171
ABSTRACT

Objective:

To establish a method for rapid determination of iodine in whole blood by direct alkali dilution inductively coupled plasma mass spectrometry (ICP-MS).

Methods:

Totally 0.50 ml whole blood sample was collected, and 2% ammonia and 0.01% Triton X-100 solution were added to constitute a total volume of 10.0 ml. After shaking to uniformity, 1.0 μg/ml rhodium and 20% isopropanol were used as on-line internal standard solution. The flow ratio of internal standard solution to the solution to be measured was 1∶16. The sample was quantitatively determined by ICP-MS. The linear range, limit of detection (LOD), accuracy and precision of the method were evaluated.

Results:

Iodine in whole blood could be determined and had a good linear relationship within the range of 0-200 μg/L, with correlation coefficient ( r) > 0.999. The LOD of the method was 0.1 μg/L, the limit of quantitation (LOQ) was 0.3 μg/L, the recovery rate of iodine in whole blood was 88.5%-106.1%, and the relative standard deviation was 2.2%-4.7% ( n=7).

Conclusions:

A method for rapid determination of iodine in whole blood by direct alkali dilution ICP-MS is successfully established. This method is accurate, simple, rapid, and highly automatic, and it can be widely applied in determination of iodine in whole blood.
Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Endemiology Year: 2020 Type: Article

Similar

MEDLINE

...
LILACS

LIS