Your browser doesn't support javascript.
loading
Comparison of fluoroscopy exposure between HIS bundle pacing and conventional pacing during procedures / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 877-881, 2020.
Article in Chinese | WPRIM | ID: wpr-868535
ABSTRACT

Objective:

To compare the differences of fluoroscopy time and dose between HIS bundle pacing and right ventricle apex pacing.

Methods:

This study includes thirty patients undergoing HIS bundle pacing (HIS group) and 32 patients undergoing right ventricular apex pacing (RVA group). The fluoroscopy time and cumulative dose (CD) to patients during surgery were recorded and analyzed.

Results:

The operation time for patients in HIS group and RVA group were (76.8±13.1) and (66.0±10.8) min ( t=3.386, P<0.001), respectively. The fluoroscopy time was (698.2±113.7) and (293.3±63.9) s ( t=14.709, P<0.001) and the CD were (391.3±70.0) and (162.3±40.5) mGy ( t=13.694, P<0.001) in HBP group and RVA group, respectively. In comparison, the fluoroscopy time and CD for HIS bundle electrode implantation were (501.2±112.3) s and (279.9±65.0) mGy, respectively, significantly higher than in the case of RVA, where the values were (103.4±30.6) s and (57.3±13.8) mGy ( t=15.864, Z=-6.524, P<0.001).

Conclusions:

Compared with right ventricular apical pacing, the HIS bundle pacing takes longer operation time, leading to higher radiation dose, which should be prudently selected.
Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2020 Type: Article

Similar

MEDLINE

...
LILACS

LIS