Your browser doesn't support javascript.
loading
A study of an independent dose verification software for brachytherapy / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6): 131-135, 2020.
Article in Chinese | WPRIM | ID: wpr-868564
ABSTRACT
Objective To report an implementation method and results of an independent brachytherapy dose verification software (DVS).Methods The DVS was developed based on Visual C++ and the modular structure design was adopted.The DICOM RT files exported from the treatment planning system (TPS) were automatically loaded into the DVS.The TG-43 formalism was employed for dose calculation.Six cervical cancer patients who underwent brachytherapy were retrospectively selected to test the DVS.Different applicators were utilized for each patient.Dosimetric parameters and γ analysis (0.1 cm,5%) were used to evaluate the dose difference between the DVS and the TPS.Results Compared with the TPS dose,the γ pass rates of the doses calculated by the DVS were higher than 98%.For CTV,the dosimetric differences were less than 0.29% and 0.53% for D100% and D90%.For bladder,rectum and sigmoid,the agreement of D0.1cm3,D1cm3 and D2cm3 within a 0.5% level.Conclusion With minimal human-computer interactions,the DVS can verify the accuracy of dose calculated by TPS for brachytherapy.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2020 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2020 Type: Article