A multicenter study on the establishment and validation of autoverification rules for coagulation tests / 中华检验医学杂志
Chinese Journal of Laboratory Medicine ; (12): 802-811, 2020.
Article in Chinese | WPRIM | ID: wpr-871978
ABSTRACT

Objective:

To establish autoverification rules for coagulation tests in multicenter cooperative units, in order to reduce workload for manual review of suspected results and shorten turnaround time (TAT) of test reports, while ensure the accuracy of results.

Methods:

A total of 14 394 blood samples were collected from fourteen hospitals during December 2019 to March 2020. These samples included Rules Establishment Group 11 230 cases, including 1 182 cases for Delta check rules; Rules Validation Group 3 164 cases, including 487cases for Delta check; Clinical Application Trial Group 77 269 cases. Samples were analyzed for coagulation tests using Sysmex CS series automatic coagulation analyzers, and the clinical information, instrument parameters, test results, clinical diagnosis, medication history of anticoagulant and other relative results such as HCT, TG, TBIL, DBIL were summarized; on the basis of historical data, the 2.5 and 97.5 percentile of all data arranged from low to high were initially accumulated; on the basis of clinical suggestions, critical values and specific drug use as well as relative guidelines, autoverification rules and limits were established.The rules were then input into middleware, in which Stage I/Stage II validation was done. Positive coincidence, negative coincidence, false negative, false positive, autoverification pass rate, passing accuracy (coincidence of autoverification and manual verification) were calculated. Autoverification rules underwent trial application in coagulation results reports.

Results:

(1) The autoverification algorisms involve 33 rules regarding PT/INR, APTT, FBG, D-dimer, FDP,Delta check, reaction curve and sample abnormalities; (2)Autoverification Establishment Group showed autoverification pass rate was 68.42% (7 684/11 230), the false negative rate was 0%(0/11230), coincidence of autoverification and manual verification was 98.51%(11 063/11 230), in which positive coincidence and negative coincidence were respectively 30.09% (3 379/11 230) and 68.42%(7 684/11 230); Autoverification Validation Group showed autoverification pass rate was 60.37%(1 910/3 164), the false negative rate was 0%(0/11 230), coincidence of autoverification and manual verification was 97.79%(3 094/3 164), in which positive coincidence and negative coincidence were respectively 37.42%(1 184/3 164) and 60.37%(1 910/3 164); (3) Trialed implementation of these autoverification rules on 77 269 coagulation samples showed that the average TAT shortened by 8.5 min-83.1 min.

Conclusions:

This study established 33 autoverification rules in coagulation tests. Validation showedthese rules could ensure test quality while shortening TAT and lighten manual workload.
Full text: Available Index: WPRIM (Western Pacific) Type of study: Controlled clinical trial / Practice guideline Language: Chinese Journal: Chinese Journal of Laboratory Medicine Year: 2020 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Controlled clinical trial / Practice guideline Language: Chinese Journal: Chinese Journal of Laboratory Medicine Year: 2020 Type: Article