Your browser doesn't support javascript.
loading
A Comparison of Central Corneal Thickness Measurements and Measurement Repeatability Using Three Imaging Modalities
Journal of the Korean Ophthalmological Society ; : 184-192, 2021.
Article in Korean | WPRIM | ID: wpr-875060
ABSTRACT
Purpose@#To compare and analyze central corneal thickness (CCT) measurements and measurement repeatability according to the corneal thickness obtained with anterior segment optical coherence tomography (AS-OCT), ultrasound pachymetry (USP), and dual Scheimpflug imaging (Galilei G4®). @*Methods@#CCT measurements obtained using AS-OCT, USP, and a dual rotating Scheimpflug camera in 120 eyes of 60 healthy subjects were compared and the repeatability was analyzed. @*Results@#The average CCT measurements using AS-OCT, USP, and a dual rotating Scheimpflug camera were 511.23 ± 37.16, 525.78 ± 39.22, and 531.43 ± 41.57 μm, respectively. The CCT measurement using dual rotating Scheimpflug imaging was statistically significantly thicker than that measured using AS-OCT (p 0.900, p < 0.001). All three methods showed high repeatability (intraclass correlation coefficient ≥ 0.996) for intra-examiner CCT measurements. @*Conclusions@#The CCT measurement results using the three methods were significantly correlated and showed high repeatability. However, the AS-OCT measurement was significantly thinner than that obtained using a dual rotating Scheimpflug camera. CCT measurement with a dual rotating Scheimpflug camera is a better alternative for USP than AS-OCT. Therefore, CCT should be interpreted in the context of the instrument used.
Full text: Available Index: WPRIM (Western Pacific) Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2021 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2021 Type: Article