ABSTRACT
OBJECTIVE@#To establish quantitative surface plasmon resonance (SPR) assay for antibodies against human platelet antigen-1a (HPA-1a).@*METHODS@#Recombinant protein was fixed on the chip surface by amino coupling method. SPR assay was used to detect the standard antibodies against HPA-1a at different conceatration. The optimal experimental parameters were determined, and standard curves were constructed with linear regression. Moreover, the sensitivity, specificity, accuracy and precision of the assay were evaluated.@*RESULTS@#The quantitative SPR assay for HPA-1a antibodies was established. The determination ranges were 0-20 IU, with accuracy (recovery rate) was 97.75%-103.08%. The intra-assay precision [coefficients of variation (CV)] was 3.53%-4.29%, and the inter-assay precision (CV) was 2.08%-4.40%. For specificity test, several kinds of monoclonal and human antibodies against platelet membrane protein were tested and no positive result was observed.@*CONCLUSION@#The established quantitative SPR assay for HPA-1a antibodies shows good sensitivity, specificity, accuracy and precision, and this rapid and simple method provides a new reference method for scientific research and clinical antibody detection.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Blood Platelets
/
Antigens, Human Platelet
/
Surface Plasmon Resonance
/
Isoantibodies
Limits:
Humans
Language:
Chinese
Journal:
Journal of Experimental Hematology
Year:
2021
Type:
Article
Similar
MEDLINE
...
LILACS
LIS