Your browser doesn't support javascript.
loading
Research of Methods to Reduce Alarm Fatigue of Monitoring System / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 481-486, 2020.
Article in Chinese | WPRIM | ID: wpr-880395
ABSTRACT
OBJECTIVE@#In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms.@*METHODS@#A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study.@*RESULTS@#The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724.@*CONCLUSIONS@#To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Arrhythmias, Cardiac / Clinical Alarms / Alert Fatigue, Health Personnel / Intensive Care Units / Monitoring, Physiologic Type of study: Controlled clinical trial Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2020 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Arrhythmias, Cardiac / Clinical Alarms / Alert Fatigue, Health Personnel / Intensive Care Units / Monitoring, Physiologic Type of study: Controlled clinical trial Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2020 Type: Article