Detecting sulfur dioxide in workplace air by ion chromatography coupled with solid adsorption tube sampling / 中国职业医学
China Occupational Medicine ; (6): 95-98, 2018.
Article in Chinese | WPRIM | ID: wpr-881672
ABSTRACT

OBJECTIVE:

To establish a detecting method for sulfur dioxide in workplace air by molecular sieve solid adsorption tube sampling and ion chromatography.

METHODS:

Air samples were collected by molecular sieve solid adsorption tubes,desorbed by distilled water,oxidized by hydrogen peroxide in weak base system,separated by anion exchange chromatography and detected by conductivity detector.

RESULTS:

The good linearity range of sulfur dioxide was0. 10-16. 00 mg/L,and the correlation coefficient was 0. 999 8. The detection limit was 0. 02 mg/L,the minimum detectable concentration was 0. 01 mg/m3. The average desorption efficiency was 96. 53%-99. 35%. The within-run and between-run relative standard deviations were 1. 73%-3. 65% and 1. 80%-4. 46% respectively. The samples could be stored at room temperature for at least 14 days.

CONCLUSION:

This method is suitable for detecting sulfur dioxide in workplace air.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Occupational Medicine Year: 2018 Type: Article

Similar

MEDLINE

...
LILACS

LIS