Your browser doesn't support javascript.
loading
Detecting 1,1,1-trichloroethane and 1,1,2-trichloroethane in workplace air by solvent desorption-gas chromatography / 中国职业医学
China Occupational Medicine ; (6): 225-232, 2019.
Article in Chinese | WPRIM | ID: wpr-881784
ABSTRACT

OBJECTIVE:

To establish a method for simultaneous determination of 1,1,1-trichloroethane(TCA) and 1,1,2-TCA in the workplace air by solvent desorption-gas chromatography.

METHODS:

The 1,1,1-TCA and 1,1,2-TCA in the workplace air were collected by activated carbon tube and desorbed with carbon disulfide. They were separated with DB-1(100.0% dimethyl polysiloxane) capillary column and detected by flame ionization detector.

RESULTS:

The good linear ranges of 1,1,1-TCA and 1,1,2-TCA were 1.340-1.338×10~4 and 1.440-1.442×10~4 mg/L, respectively. The correlation coefficients were greater than 0.999, with the detection limits of 0.100 and 0.140 mg/L, respectively. The average desorption efficiencies of 1,1,1-TCA and 1,1,2-TCA were 98.6%-99.6% and 94.9%-96.2%. The within-run and between-run relative standard deviations were 0.4%-0.9% and 0.5%-1.1%, respectively. The samples could be stored at room temperature for at least 14 days.

CONCLUSION:

This method has high desorption efficiency and sensitivity, good precision and simple operation, which is suitable for simultaneous detection of 1,1,1-TCA and 1,1,2-TCA in the workplace air.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Occupational Medicine Year: 2019 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Occupational Medicine Year: 2019 Type: Article