Your browser doesn't support javascript.
loading
Determination of volatile organic compounds in air using portable gas chromatography-mass spectrometry / 中国职业医学
China Occupational Medicine ; (6): 85-88, 2021.
Article in Chinese | WPRIM | ID: wpr-881976
ABSTRACT

OBJECTIVE:

To explore a method for detecting volatile organic compounds(VOCs) in air by a portable gas chromatography-mass spectrometer(GC-MS).

METHODS:

A portable GC-MS hand-held probe was used for sampling and detection. All effluent components were qualitatively analyzed by the standard spectral library of the United State National Institute of Standards and Technology. The percentage of peak area of the component was calculated by normalization method. The static distribution method was adopted in the semi-quantitative and quantitative analysis using nitrogen as the diluting gas to prepare different mass concentration of 13 kinds of VOCs in the mixed standard gas. The retention time and characteristic ions were used for qualitative analysis, and the quantitated full scanning mode was used for quantitative analysis.

RESULTS:

The minimum detected mass concentration of the 13 chemical harmful factors was 0.02-0.10 mg/m~3, and the minimum quantitative mass concentration was 0.07-0.38 mg/m~3. The recovery rate was 84.76%-116.56%, and the within-and between-run relative standard deviations were 4.10%-12.50% and 8.17%-14.36%, respectively.

CONCLUSION:

The portable GC-MS instrument could be used for qualitative alalysis, peak area percentage determination, semi-quantitative and quantitative analysis of VOCs in the workplace air or sudden chemical poisoning accidents.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Qualitative research Language: Chinese Journal: China Occupational Medicine Year: 2021 Type: Article

Similar

MEDLINE

...
LILACS

LIS