Design and Verification of Isokinetic Muscle Strength Rehabilitation Training Equipment Based on Oil Resistance / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 390-393, 2021.
Article in Chinese | WPRIM | ID: wpr-888630
ABSTRACT
This study designed an isokinetic muscle strength rehabilitation training equipment, based on oil resistance. The equipment detected joint angles during movement through displacement sensors to measure the angular velocity in real-time. Then it can achieve flow resistance adjustment and constant velocity movement, based on the flow area of the proportional valve calculated by PID algorithm and the deviation between the set speed and the measured speed. The equipment can also collect muscle strength during exercise through force-sensitive sensors to achieve real-time and accurate assessment of muscle strength. Moreover, the study preliminarily certificated the isokinetic motion the results showed that the mean error of peak torque test is less than 5%, and the correlation is between 0.89 and 0.92 (
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Muscle, Skeletal / Torque / Exercise Therapy / Muscle Strength / Movement Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2021 Type: Article