Fast Separation Method of Continuous Recording Data in Dynamic Electrocardiogram Database / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 442-445, 2021.
Article in Chinese | WPRIM | ID: wpr-888642
ABSTRACT
The square waves and pulses with obvious porter characteristics are inserted into each different adjacent existing dynamic electrocardiogram data to solve the time-consuming problem in the currently used manual input verification method. The standard database files are converted into analog siginals, then output automatically to the ECG acquisition device according to the sequence of the database files in one time. The data recorded in the acquisition equipment is separated according to the interval data protocol, so as to achieve the purpose of rapid digital examination of dynamic electrocardiogram.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Electrocardiography / Heart Rate Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2021 Type: Article