Accuracy of Pedicle Screw Insertion Using Fluoroscopy-Based Navigation-Assisted Surgery : Computed Tomography Postoperative Assessment in 96 Consecutive Patients
Journal of Korean Neurosurgical Society ; : 16-20, 2014.
Article in English | WPRIM | ID: wpr-89974
ABSTRACT

OBJECTIVE:

Two-dimensional fluoroscopy-based computerized navigation for the placement of pedicle screws offers the advantage of using stored patient-specific imaging data in providing real-time guidance during screw placement. The study aimed to describe the accuracy and reliability of a fluoroscopy-based navigation system for pedicle screw insertion.

METHODS:

A total of 477 pedicle screws were inserted in the lower back of 96 consecutive patients between October 2007 and June 2012 using fluoroscopy-based computer-assisted surgery. The accuracy of screw placement was evaluated using a sophisticated computed tomography protocol.

RESULTS:

Of the 477 pedicle screws, 461 (96.7%) were judged to be inserted correctly. Frank screw misplacement [16 screws (3.3%)] was observed in 15 patients. Of these, 8 were classified as minimally misplaced (4 mm). No complications, including nerve root injury, cerebrospinal fluid leakage, or internal organ injury, were observed in any of the patients.

CONCLUSION:

The accuracy of pedicle screw placement using a fluoroscopy-based computer navigation system was observed to be superior to that obtained with conventional techniques.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Cerebrospinal Fluid / Surgery, Computer-Assisted Limits: Humans Language: English Journal: Journal of Korean Neurosurgical Society Year: 2014 Type: Article