Your browser doesn't support javascript.
loading
Determination of digoxin in human plasma by LC-MS/MS and its application in pediatric patients / 中国药科大学学报
Journal of China Pharmaceutical University ; (6): 719-724, 2021.
Article in Chinese | WPRIM | ID: wpr-906766
ABSTRACT
@#The aim of the study was to develop a simple, rapid and accurate LC-MS/MS method for the determination of digoxin.Digoxin-d3 was taken as the internal standard (IS), and sample preparation was achieved by liquid-liquid extraction.Chromatographic separation was performed on a Kinetex C18 column (2.1 mm × 50 mm, 2.6 μm; Phenomenex) using an isocratic elution with merely 2 min for each sample.The mobile phase consisted of water and acetonitrile solutions, both containing 1 mmol/L ammonium acetate and 1 mmol/L formic acid (55∶45).The detection was conducted on a TripleQuadTM 4500MD mass spectrometer coupled with electrospray ionization interface under positive-ion multiple reaction monitoring mode.The transitions were m/z 798.5 → 651.3 and m/z 801.6 → 654.4 for digoxin and digoxin-d3, respectively.Results showed that the method was linear over the range of 0.100-20.0 ng/mL.The selectivity, accuracy and precision, recovery and stability of the method were all within the acceptable limits with no matrix effect.This method was successfully applied to a girl treated with digoxin with substantial improvement of therapeutic effect and elimination of toxic reaction, so it can provide valuable fuidance and reference for individualized medication in clinical practice.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Journal of China Pharmaceutical University Year: 2021 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Journal of China Pharmaceutical University Year: 2021 Type: Article