Your browser doesn't support javascript.
loading
Dose estimation of X-ray radiography based on Chinese pediatric reference phantoms / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 363-367, 2021.
Article in Chinese | WPRIM | ID: wpr-910322
ABSTRACT

Objective:

The Chinese pediatric mesh-type reference phantoms were developed and applied in the dose assessment of X-ray radiography.

Methods:

The 5- and 10-year-old Chinese pediatric mesh-type reference phantoms were developed based on the CT data. Based on the phantoms, the Monte Carlo method was applied to simulate the X-ray radiography to calculate the organ dose conversion coefficient and the effective dose conversion coefficient in different radiography conditions.

Results:

The 5- and 10-year-old Chinese pediatric mesh-type reference phantoms were developed, and the physical parameters were consistent with the national standard. The differences of the organ mass between the established phantoms and reference data were within 2%. The database of the pediatric chest posteroanterior projection and abdominal anteroposterior projection, tube voltage 60-90 kVp, total filtration 2.5~4 mmAl were simulated, and the difference between the effective dose conversion factor and the literature result was within 3%.

Conclusions:

The established Chinese pediatric mesh-type reference phantoms can be applied in the studies of radiation protection and clinical medicine and their result can provide an important reference for the dose assessment of the pediatric X-ray radiography.