Your browser doesn't support javascript.
loading
Phantom study of the influence of time of flight and point spread function on PET/CT image quality and standardized uptake value / 中华核医学与分子影像杂志
Chinese Journal of Nuclear Medicine and Molecular Imaging ; (6): 420-424, 2021.
Article in Chinese | WPRIM | ID: wpr-910782
ABSTRACT

Objective:

To explore the influence of time of flight (TOF) and point spread function (PSF) on PET/CT image quality and standardized uptake value (SUV) based on a phantom study.

Methods:

PET/CT imaging were performed using Jaszczak phantom and International Electrotechnical Commission (IEC) body phantom respectively, and conventional clinical imaging protocol (3 min/bed) was used for three consecutive imaging. The ordered subsets expectation maximization (OSEM), OSEM+ TOF, OSEM+ PSF and OSEM+ TOF+ PSF were used to reconstruct the raw data respectively. The differences of image resolution, image uniformity, contrast, signal to noise ratio (SNR) and SUV among different reconstruction algorithms were compared by using one-way analysis of variance and the least significant difference (LSD) t test. The correlations between SNR, mean SUV (SUV mean), maximum SUV (SUV max) and the sphere diameter were analyzed by Pearson correlation analysis.

Results:

The minimum image resolution for the hot column was 6.4 mm by OSEM and OSEM+ PSF reconstruction, while it was 4.8 mm by OSEM+ TOF and OSEM+ TOF+ PSF reconstruction. The contrast was significantly better by OSEM+ TOF ((78.56±1.21)%) and OSEM+ TOF+ PSF ((78.85±1.17)%) reconstruction than that by OSEM reconstruction ((73.44±1.47)%; F=61.068, t values 9.503, 10.018, both P<0.001). The maximum and minimum nonuniformity percentages were significantly better for images reconstructed with OSEM+ PSF than those reconstructed with OSEM ( F values 10.286, 27.630, t values -2.599, 7.698, both P<0.05). In IEC phantom imaging, the SNR and SUV max of hot sphere were higher by OSEM+ PSF and OSEM+ TOF+ PSF reconstruction than those by OSEM reconstruction (SNR (98.99±34.76)%, (98.29±28.66)%, (73.64±26.05)%; F=5.712, t values 3.209, 3.412, both P<0.05; SUV max 8.96±2.72, 9.28±2.17, 8.01±2.21; F=3.931, t values 2.154, 2.863, both P<0.05), but there was no significant difference between OSEM+ TOF and OSEM reconstruction ( t values 0.297, 0.272, both P>0.05). Among the four reconstruction methods, the SNR and SUV mean increased with the increase of the sphere diameter ( r values 0.913-0.963, all P<0.05), but SUV max in images reconstructed with OSEM+ PSF and OSEM+ TOF+ PSF had no significant correlations with the sphere diameter ( r values 0.496, 0.614, both P>0.05).

Conclusions:

Under specific acquisition and reconstruction conditions, TOF mainly improves image resolution and contrast, while PSF mainly improves image uniformity and SNR. The combination of the two method can obtain better image quality and significantly improve the SUV of hot lesions.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Practice guideline Language: Chinese Journal: Chinese Journal of Nuclear Medicine and Molecular Imaging Year: 2021 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Practice guideline Language: Chinese Journal: Chinese Journal of Nuclear Medicine and Molecular Imaging Year: 2021 Type: Article