ABSTRACT

Objective:

To establish the allowable total error (TEa) of the national external quality assessment (EQA) program in line with the current quality level of serum folate measurement in China.

Methods:

The data of serum total folate test in the clinical laboratory of a hospital in Beijing in 2016 were collected, and the Stata SE 15 software was used for Monte Carlo simulation to obtain the false-negative rate under different bias and inaccuracy conditions. The Origin Pro 9.1 software was used to make the contour figure. The TEa of serum total folate test is derived based on the acceptable false-negative rate. National EQA data of serum total folate in 2020 were collected to calculate the pass rate of participating laboratories and the laboratory pass rate of quality control products at each level under the five TEa derived from the analysis performance on clinical outcomes, biological variation, and the evaluation criterion of national EQA.

Results:

Based on the influence of analytical performance on clinical outcomes, the TEa was 10%. Under this TEa, the pass rate of the first EQA program of serum total folate in 2020 was more than 80%, and the pass rate of the second time was 73.1%. Under the minimum (46.57%) and appropriate level of TEa (15.52%) derived from biological variation and national EQA evaluation criterion, the pass rate of serum total folate in the two EQA programs in 2020 exceeded 85%.

Conclusion:

The analytical performance of serum total folate in China cannot meet the requirements of TEa derived based on the effect of analytical performance on clinical outcomes. An appropriate level of TEa derived based on biological variation (15.52%) is suggested as the recommended criterion for the TEa of serum total folate test.