Design and Application of Intelligent Management System of Artificial Airway Airbag Pressure in Intensive Care / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 645-649, 2021.
Article
in Chinese
| WPRIM
| ID: wpr-922076
ABSTRACT
In order to solve the problem of continuous monitoring and automatic regulation of patient airbag pressure in intensive care unit, the study designed an intelligent management system of artificial airway airbag pressure. It can realize real-time monitoring and automatic control of airbag pressure. Its pressure data was sent to the PC in real time by the serial port. It can realize the display, store, review and analysis of pressure data. Its clinical application effect was discussed. Experiments showed that the system can monitor airbag pressure in real time and control the pressure to stabilize at 25~30 cmH