Your browser doesn't support javascript.
loading
Preparation of air sampling tube for chlorobenzene compounds and establishment of its matching detecting method / 中国职业医学
China Occupational Medicine ; (6): 425-430, 2021.
Article in Chinese | WPRIM | ID: wpr-923212
ABSTRACT

OBJECTIVE:

To prepare and develop a GDH-2 air sampling tube for detecting 12 kinds of chlorobenzenes(CBs) in workplace air and to establish a matching detecting method.

METHODS:

The self-developed GDH-2 air sampling tube was filled with ion exchange resin and activated carbon, and the mass ratio was 10 ∶1. The GDH-2 air sampling tube was used to collect 12 kinds of CBs with coexistence of gaseous and aerosol in the air. After elution with toluene, they were separated on a chromatographic column and determined by microcell electron capture detector.

RESULTS:

The quantitative detecting range of the method was 0.51×10~(-3)-6 000.00 mg/L, with the correlation coefficients greater than 0.999 4. The minimum detection concentration was 0.02-61.99 μg/m~3, and the minimum quantitative concentration was 0.05-206.62 μg/m~3. The average desorption efficiency was 90.8%-104.0%. The within-run relative standard deviation(RSD) was 1.0%-5.7%, and the between-run RSD was 3.0%-7.3%. The samples can be stored at room temperature for at least 26 days.

CONCLUSION:

The self-developed GDH-2 air sampling tube and its matching measuring method can be used for the collection and determination of the 12 kinds of CBs in the air of workplace.