Your browser doesn't support javascript.
loading
Optimization of Parallel-Hole Collimator for Small Gamma Camera Based on Pixellated Crystal Array / 의학물리
Korean Journal of Medical Physics ; : 291-297, 2008.
Article in Korean | WPRIM | ID: wpr-93128
ABSTRACT
The purpose of this study is to optimize a parallel-hole collimator for small gamma camera having the pixellated crystal array and evaluate the effect of crystal-collimator misalignment on the image quality using a simulation tool GATE (Geant4 Application for Tomographic Emission). The spatial resolution and sensitivity were measured for the various size of hexagonal-hole and matched square-hole collimators with a Tc-99m point source and the uniformity of flood image was estimated as a function of the angle between crystal array and collimator by misalignment. The results showed that the spatial resolution and sensitivity were greatly improved by using the matched collimator and the uniformity was reduced by crystal-collimator misalignment.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Gamma Cameras Language: Korean Journal: Korean Journal of Medical Physics Year: 2008 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Gamma Cameras Language: Korean Journal: Korean Journal of Medical Physics Year: 2008 Type: Article