Your browser doesn't support javascript.
loading
Fabrication and Characterization of a One-dimensional Fiber-optic Dosimeter for Electron Beam Therapy Dosimetry / 의학물리
Korean Journal of Medical Physics ; : 285-290, 2008.
Article in Korean | WPRIM | ID: wpr-93129
ABSTRACT
In this study, we have fabricated a one-dimensional fiber-optic dosimeter for electron beam therapy dosimetry. Each fiber-optic dosimeter has an organic scintillator with a plastic optical fiber and it is embedded and arrayed in the plastic phantom to measure one-dimensional high energy electron beam profile of clinical linear accelerator. The scintillating lights generated from each sensor probe are guided by plastic optical fibers to the multi-channel photodiode amplifier system. We have measured one-dimensional electron beam profiles in a PMMA phantom according to different field sizes and energies of electron beam. Also, the isodose and three-dimensional percent depth dose curves in a PMMA phantom are obtained using a one-dimensional fiber-optic dosimeter with different electron beam energies.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Particle Accelerators / Plastics / Polymethyl Methacrylate / Electrons / Optical Fibers / Light Language: Korean Journal: Korean Journal of Medical Physics Year: 2008 Type: Article