Your browser doesn't support javascript.
loading
In vitro evaluation of positioning accuracy of trephine bur at different depths by dynamic navigation / 北京大学学报(医学版)
Journal of Peking University(Health Sciences) ; (6): 146-152, 2022.
Article in Chinese | WPRIM | ID: wpr-936126
ABSTRACT
OBJECTIVE@#To evaluate the accuracy of trephine bur drilling at different depths guided by dynamic navigation system in 3D printing in vitro model.@*METHODS@#A model at the depth of 5 mm, 10 mm, and 15 mm from the outer surface of which hemispherical cavities was reserved and the 3D printing technology was used to make the standardized model with Veroclear resin. The cone beam CT (CBCT) was taken and the data were imported into the dynamic navigation software (DCARER, China) to establish navigation path programming. Under the guidance of dynamic navigation, a trephine bur with a diameter of 4.5 mm was used to complete the access operation. At each depth, 10 approaches were completed. The postoperative model CBCT was taken. The approach trajectory under navigation was reconstructed and compared with the designed path. The two-dimensional distance deviation, depth deviation, three-dimensional distance deviation, and angle deviation between the actually prepared path and the designed path were calculated.@*RESULTS@#At the depth of 5 mm, the two-dimensional distance deviation between the end position of the prepared path and the designed path was (0.37±0.06) mm, the depth deviation was (0.06±0.05) mm, the three-dimensional distance deviation was (0.38±0.07) mm, and the angle deviation was 2.46°±0.54°; At the depth of 10 mm, the four deviations between the end position of prepared path and the designed path were (0.44±0.05) mm, (0.16±0.06) mm, (0.47±0.05) mm, and 2.45°±1.21°, respectively; At the depth of 15 mm, the four deviations were (0.52±0.14) mm, (0.16±0.07) mm, (0.55±0.15) mm, and 3.25°±1.22°, respectively. With the increase of entry depth, the three-dimensional and depth accuracy of dynamic navigation system decreased (P < 0.01), and the positioning angle deviation had no relation with the entry depth (P>0.01).@*CONCLUSION@#Dynamic navigation technology can achieve high positioning accuracy in the depth range of 15 mm, but its deviation increases with the increase of entry depth.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: China / Surgery, Computer-Assisted / Cone-Beam Computed Tomography / Printing, Three-Dimensional Country/Region as subject: Asia Language: Chinese Journal: Journal of Peking University(Health Sciences) Year: 2022 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: China / Surgery, Computer-Assisted / Cone-Beam Computed Tomography / Printing, Three-Dimensional Country/Region as subject: Asia Language: Chinese Journal: Journal of Peking University(Health Sciences) Year: 2022 Type: Article