A phantom study of the effect of deviation from isocentric points on CT image quality / 中华放射学杂志
Chinese Journal of Radiology ; (12): 1237-1241, 2022.
Article in Chinese | WPRIM | ID: wpr-956781
ABSTRACT

Objective:

To investigate the effect of deviation from the isocenter point on the quality of CT images at the same radiation dose.

Methods:

A 160-layer CT scanner was used to scan the phantom at isocenter and deviations of 3, 6, 9, 12 and 15 cm. CT was performed with the following parameters 120 kVp; 400 mAs; slice thickness, 1 mm; and slice increment, 1 mm. Images were reconstructed using the filtered back projection algorithm. Noise power spectrum (NPS), task transfer function (TTF) and detectability index (d′) were measured. NPS peak was used to quantify the noise magnitude and TTF 50% was used to quantify the spatial resolution. NPS, TTF and d′ were compared using one-way ANOVA.

Results:

The NPS average spatial frequency, spatial resolution and d′ values gradually decreased as the offset distance increased and the amount of noise increased. NPS peak at isocenter and deviations of 3 cm, 6 cm, 9 cm, 12 cm and 15 cm were (94.31±1.48), (104.25±1.46), (131.44±1.96), (171.86±1.91), (224.05±1.37), (286.51±2.09)HU 2·mm 2, respectively ( F=37 241.91, P<0.001). And d′ values of 2 mm low-contrast lesions were 3.51±0.06, 3.31±0.04, 3.01±0.04, 2.59±0.06, 2.21±0.03, 1.88±0.03, respectively. The reduction in spatial resolution was more pronounced for high contrast, and the d′ values decreased to a similar extent for various types of lesions. The noise was increased by about 82%, the high contrast spatial resolution was decreased by about 12%, and the d′ value was decreased by about 26% at 9 cm from the isocenter point (all P<0.05). The noise was increased by about 204%, the high contrast spatial resolution was decreased by about 27%, and the d′ value was decreased by about 45% at 15 cm from the isocenter (all P<0.05).

Conclusion:

The CT image quality was decreased with the increase of the offset distance from the CT isocenter point. The image quality was severely compromised at offset distances greater than 9 cm.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiology Year: 2022 Type: Article