Your browser doesn't support javascript.
loading
Analysis of different fixation devices in precision radiotherapy for prostate cancer under bowel and bladder preparation protocol / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6): 716-721, 2022.
Article in Chinese | WPRIM | ID: wpr-956901
ABSTRACT

Objective:

To provide evidence for the selection of fixation devices and CTV to PTV margins (M ptv) in precision radiotherapy for pelvic tumors by analyzing three fixation devices in precision radiotherapy for prostate cancer.

Methods:

From April 2015 to December 2020, 133 prostate cancer patients treated with pelvic drainage area irradiation in our center were retrospectively analyzed. The patients were fixed with 1.2m vacuum bag (n=39), 1.8m vacuum bag (n=44) and personalized prone plate by our center (n=50). Each patient was asked to complete our bowel and bladder preparation process before positioning and radiotherapy. The registration of CBCT to planned CT before each treatment adopted the same registration box and algorithm. Setup errors in the SI, LR and AP directions under qualified bowel and bladder conditions were recorded. Setup errors in three directions under three fixation devices and corresponding M ptv values were analyzed. The correlation between setup errors with age and body mass index (BMI) was analyzed.

Results:

Analysis of 3333 setup errors data showed in the SI and LR directions, the mean setup errors of 1.2m vacuum bag (3.26mm, 2.34mm) were greater than those of 1.8m vacuum bag (2.51mm, P<0.001; 1.90mm, P<0.001), and personalized prone plate (3.07mm, P=0.066; 2.10 mm, P=0.009). In the AP direction, the mean setup errors of 1.2m vacuum bag (supine)(2.20mm) were smaller than those of 1.8m vacuum bag (3.33mm, P<0.001) and personalized prone plate (3.61mm, P<0.001). The setup errors of 1.8m vacuum bag in all directions were smaller than those of personalized prone plate (P≤0.028). According to Van Herk's expansion formula, the M ptv of 1.2m vacuum bag in three directions was approximately 4 mm. The M ptv of 1.8m vacuum bag and personalized prone plate in the SI and LR directions was approximately 3 mm, and more than 5 mm in the AP direction. The setup errors were not correlated with age or BMI.

Conclusions:

From the setup errors results of three devices, 1.8m vacuum bag is the best, followed by personalized prone plate. And supine position is better than prone position in the AP direction.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2022 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2022 Type: Article