Your browser doesn't support javascript.
loading
Discussion on the evaluation of radiation protection of “four-in-one” dental X-ray equipment / 中国辐射卫生
Chinese Journal of Radiological Health ; (6): 322-327, 2023.
Article in Chinese | WPRIM | ID: wpr-978437
ABSTRACT
Objective To evaluate the radiation protection of “four-in-one” dental X-ray equipment and to investigate the safety interlock of the equipment by measuring the scattered radiation at the position of the patient during operation. Methods A cone-beam CT dental phantom was used to simulate the patient’s head. The intra-oral and extra-oral components of the “four-in-one” X-ray equipment were installed in a 5 m2 room. The scattered radiation at patient position was measured using a γ/X-ray survey meter, and the effects of intra-oral and extra-oral components were compared. Results For a 5 m2 room, when CBCT was exposed under typical conditions, the dose at the patient's position was 10.70 uSv/h when there was an intra-oral component and 10.60 uSv/h when there was no intraoral component. The intra-oral part did not affect the radiation dose at the patient's position. When the intra-oral component was exposed, the dose rate at the patient's position was 4.05-6.85 uSv/h, and the extra-oral part did not affect the scattered dose of the patient examined with intra-oral components. Conclusion The evaluation of radiation protection of new equipment must comprehensively consider radiation safety and equipment operation safety. The results of this study provide suggestions for clinical radiation protection supervision and evaluation of “four-in-one” dental X-ray equipment.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Health Year: 2023 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Health Year: 2023 Type: Article