Stripping of the Hexagonal Recess in the Process of LCP (Locking Compression Plate) Removal
Journal of the Korean Fracture Society ; : 283-287, 2006.
Article in Korean | WPRIM | ID: wpr-9953
ABSTRACT

PURPOSE:

To report the difficulties in the process of locking head screw removal due to the stripping of the hexagonal recess of the screw head. MATERIALS AND

METHODS:

We have removed 113 5.0-self tapping locking head screws and 202 3.5-self tapping locking head screws from 34 patients with fracture healing and 5 patients complicated with infection. All of the operations were done by one surgeon. All the screws were placed with the use of torque limiting attachment or driver.

RESULTS:

All of 113 5.0-self tapping locking head screws were removed without difficulties with an usual manner. 21 out of 202 3.5-self tapping locking head screws were removed with many difficulties due to the stripping of the hexagonal recess. 3 screws were removed successfully with the use of conical extraction screw. 12 screws were taken out by further stripping and destruction of the screw head. In 6 situations where the only one screw was left stripped, the plate was bent around the stripped screw and then it was removed by turning the plate as a handle. One screw was removed with the partial breakage of the near cortex upon lifting the plate after failed attempt of using conical extraction screw.

CONCLUSION:

Although we have followed the guidelines at the time of insertion we have experienced difficulties in the removal of 3.5 locking head screws due to the stripping of the hexagonal recess. Care should be taken at the time of removal of the locking plate especially for the 3.5 locking screws.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Fracture Healing / Lifting / Torque / Head Limits: Humans Language: Korean Journal: Journal of the Korean Fracture Society Year: 2006 Type: Article