Your browser doesn't support javascript.
loading
Comparación de la precisión de la longitud de trabajo de cuatro localizadores apicales electrónicos con limas de acero inoxidable y níquel-titanio en conductos mesiovestibulares de primeras molares inferiores / Comparison of the precision of the working length of four electronic apical locators with stainless steel files and titanium nickel in mesiovestibular ducts of first lower molars
Mio-Suclupe, Rosa Eudocia; García-Rupaya, Carmen Rosa.
Afiliación
  • Mio-Suclupe, Rosa Eudocia; Universidad Científica del Sur. Lima. PE
  • García-Rupaya, Carmen Rosa; Universidad Científica del Sur. Lima. PE
Rev. cient. odontol ; 8(3): e28, sept.-dic. 2020. tab
Article en Es | LILACS, LIPECS | ID: biblio-1253434
Biblioteca responsable: PE1.1
RESUMEN

Objetivo:

El propósito del estudio in vitro fue comparar la precisión de la longitud de trabajo de Root ZX mini, Proper Pixi, Woodpex III y Raypex 6 utilizando limas de acero inoxidable VDW C-Pilot n.º 15 y de níquel-titanio Maillefer n.º15. Materiales y

métodos:

Este estudio fue experimental in vitro. La muestra fue de treinta y cinco conductos mesiovestibulares de primeras molares inferiores, con cuatro localizadores apicales electrónicos Root ZX mini, Proper Pixi, Woodpex III y Raypex6. La longitud real se determinó con un calibrador Vernier digital y la medición electrónica se realizó utilizando limas de acero inoxidable y de níquel-titanio.

Resultados:

Al comparar la longitud real y longitud electrónica, no se encontró diferencias estadísticamente significativas (p < 0,05).

Conclusiones:

Se halló diferencias estadísticamente significativas al comparar la precisión de Root ZX mini, Proper Pixi, Woodpex III y Raypex 6 con limas de acero inoxidable; mientras que la precisión con limas de níquel-titanio no se vio afectada. (AU)
ABSTRACT

Objective:

The purpose of this study was to compare the precision of the working length of Root ZX mini, Proper Pixi, Woodpex III and Raypex 6 using VDW C-Pilot No. 15 with Maillefer nickel titanium No. 15 and stainless steel files in mesiovestibular ducts of first lower molars. Materials and

methods:

This was an experimental in vitro study, including 35 mesiobuccal canals of lower first molars. The real length of 4 electronic apical locators, Root ZX mini, Proper Pixi, Woodpex III and Raypex6, was determined with a digital Vernier caliper, and electronic measurements were recorded using stainless steel and nickel titanium files.

Results:

There were no statistically significant differences between the real and the electronic lengths of the different tools studied (p <0.05).

Conclusions:

Statistically significant differences were obtained when comparing the precision of Root ZX mini, Proper Pixi, Woodpex III and Raypex 6 with stainless steel files, while the precision of nickel titanium files was not affected. (AU)
Asunto(s)
Palabras clave

Texto completo: 1 Índice: LILACS Asunto principal: Titanio / Materiales Dentales / Diente Molar / Níquel Límite: Humans Idioma: Es Revista: Rev. cient. odontol Asunto de la revista: Odontologia Año: 2020 Tipo del documento: Article

Texto completo: 1 Índice: LILACS Asunto principal: Titanio / Materiales Dentales / Diente Molar / Níquel Límite: Humans Idioma: Es Revista: Rev. cient. odontol Asunto de la revista: Odontologia Año: 2020 Tipo del documento: Article