Intra and intersession repeatability and reliability of dynamic parameters in pressure platform assessments on subjects with simulated leg length discrepancy. A cross-sectional research
Pereiro-Buceta, Héctor; Calvo-Lobo, César; Becerro-de-Bengoa-Vallejo, Ricardo; Losa-Iglesias, Marta Elena; Romero-Morales, Carlos; López-López, Daniel; Martínez-Jiménez, Eva-María.
Afiliación
  • Pereiro-Buceta, Héctor; Universidade da Coruña. Department of Health Sciences, Faculty of Nursing and Podiatry. Health and Podiatry Group. Ferrol. ES
  • Calvo-Lobo, César; Universidad Complutense de Madrid. Facultad de Enfermería, Fisioterapia y Podología. Madrid. ES
  • Becerro-de-Bengoa-Vallejo, Ricardo; Universidad Complutense de Madrid. Facultad de Enfermería, Fisioterapia y Podología. Madrid. ES
  • Losa-Iglesias, Marta Elena; Universidad Rey Juan Carlos. Faculty of Health Sciences. Alcorcón. ES
  • Romero-Morales, Carlos; Universidad Europea de Madrid, Villaviciosa de Odón. Faculty of Sport Sciences. Madrid. ES
  • López-López, Daniel; Universidade da Coruña. Department of Health Sciences. Health and Podiatry Group. Ferrol. ES
  • Martínez-Jiménez, Eva-María; Universidad Complutense de Madrid. Facultad de Enfermería, Fisioterapia y Podología. Madrid. ES
São Paulo med. j ; São Paulo med. j;139(5): 424-434, May 2021. tab, graf
Article en En | LILACS | ID: biblio-1290254
Biblioteca responsable: BR1.1
ABSTRACT
ABSTRACT

BACKGROUND:

Leg length discrepancy (LLD) may play a key role in exercise biomechanics. Although the Podoprint platform has been used in dynamic pressure studies, there are no data regarding the reliability and repeatability of dynamic measurements under simulated LLD conditions.

OBJECTIVES:

To determine the intra and intersession repeatability and reliability of dynamic parameters of the Podoprint pressure platform under simulated LLD conditions. DESIGN AND

SETTING:

Observational cross-sectional study at a public university.

METHODS:

Thirty-seven healthy volunteers participated in this study. LLD was simulated using ethyl vinyl acetate plantar lifts with heights of 5 mm, 10 mm, 15 mm and 20 mm located under the right shoe of each volunteer. The procedure was performed to capture the dynamic parameters of each participant under five different simulated LLD conditions. Stance time, mean pressure and peak pressure measurements were registered in three trials for each foot and each LLD level. Data were collected during two separate testing sessions, in order to establish intrasession and intersession reliability.

RESULTS:

The intraclass correlation coefficients (ICCs) for intrasession reliability ranged from 0.775 to 0.983 in the first session and from 0.860 to 0.985 in the second session. The ICCs for intersession reliability ranged from 0.909 to 0.990. Bland-Altman plots showed absence of systematic measurement errors.

CONCLUSIONS:

The results from this study indicate that the Podoprint platform is a reliable system for assessing dynamic parameters under simulated LLD conditions. Future studies should evaluate plantar pressures under LLD conditions, in association with exercise, biomechanics and musculoskeletal disorders.
Asunto(s)
Palabras clave

Texto completo: 1 Índice: LILACS Asunto principal: Pie / Pierna Tipo de estudio: Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: São Paulo med. j Asunto de la revista: Cirurgia Geral / Ciˆncia / Ginecologia / MEDICINA / Medicina Interna / Obstetr¡cia / Pediatria / Sa£de Mental / Sa£de P£blica Año: 2021 Tipo del documento: Article

Texto completo: 1 Índice: LILACS Asunto principal: Pie / Pierna Tipo de estudio: Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: São Paulo med. j Asunto de la revista: Cirurgia Geral / Ciˆncia / Ginecologia / MEDICINA / Medicina Interna / Obstetr¡cia / Pediatria / Sa£de Mental / Sa£de P£blica Año: 2021 Tipo del documento: Article