Una mirada a la seguridad del ultrasonido en ecotomografía obstétrica / A look at the safety of ultrasound in obstetric echotomography
Jeldes-Maturana, Jorge; Ramírez-Peña, Rodrigo; Zúñiga-Clavijo, Patricia; Ogalde-Baeza, Sebastián; Miño-Cornejo, José S; Fuentes-Alburquenque, Mauricio; Pastén-Avalos, Miriam; Canals-Lambarri, Mauricio.
  • Jeldes-Maturana, Jorge; Universidad de Valparaíso. Escuela de Tecnología Médica. Valparaíso. CL
  • Ramírez-Peña, Rodrigo; Universidad de Valparaíso. Escuela de Tecnología Médica. Valparaíso. CL
  • Zúñiga-Clavijo, Patricia; Universidad de Valparaíso. Escuela de Tecnología Médica. Valparaiso. CL
  • Ogalde-Baeza, Sebastián; Clínica Espacio EME. Servicio de Ecografía. Santiago. CL
  • Miño-Cornejo, José S; Clínica Espacio EME. Servicio de Ecografía. Santiago. CL
  • Fuentes-Alburquenque, Mauricio; Universidad de Chile, Santiago. Facultad de Medicina. Escuela de Salud Pública. Santiago. CL
  • Pastén-Avalos, Miriam; Universidad Andrés Bello. Facultad de Medicina. Escuela de Tecnología Médica. Santiago. CL
  • Canals-Lambarri, Mauricio; Universidad de Chile. Facultad de Medicina. Escuela de Salud Pública. Santiago. CL
Rev. chil. obstet. ginecol. (En línea) ; 87(4): 279-284, ago. 2022. tab, graf
Artículo en Español | LILACS | ID: biblio-1407854
RESUMEN
Resumen Los índices térmico y mecánico son estimadores de riesgo fetal en una ecografía y se deben controlar sus valores minimizando la exposición. Comparamos los valores de los índices térmico y mecánico obtenidos en exámenes ecográficos obstétricos de pacientes gestantes, con el valor recomendado por la comunidad internacional. Se realizó la estimación de estos índices en 421 ecografías obstétricas en 2019. Los valores fueron comparados entre ellos y con el valor recomendado para cada índice y según el modo de visualización (B, Doppler color y Doppler espectral). Del total de la muestra, para el índice térmico en modo Doppler color un 0,24% superó el valor estándar recomendado y en modo Doppler espectral un 2,85%. Para el índice mecánico se sobrepasó el valor recomendado en modo B en un 11,16%, en un 8,08% en modo Doppler color y, por último, en un 0,48% para el modo Doppler espectral. Los índices mecánico y térmico en esta muestra de exámenes ecográficos obstétricos se encuentran en promedio bajo el valor de referencia. Sin embargo, existe un número importante de casos en que se superan las normas, lo que debe ser una voz de alerta para la comunidad médica.
ABSTRACT
Abstract The thermal and mechanical indices are the best estimators of fetal risk in an ultrasound and their values should be controlled in order to minimize exposure as much as possible. We compared the values of the thermal and mechanical indices obtained in obstetric ultrasound examinations of pregnant patients, with the value recommended by the international community. These indices were estimated in 421 obstetric ultrasounds during 2019. The estimated values were compared with each other and with the recommended value for each index and according to the display mode (B, color Doppler and spectral Doppler). Of the total sample, for the thermal index in color Doppler mode, 0.24% exceeded the recommended standard value and 2.85%. in spectral Doppler mode. For the mechanical index, the recommended value was exceeded in B mode by 11.16%, in 8.08% in color Doppler mode and, finally, by 0.48% for spectral Doppler mode. The mechanical and thermal indices in this sample of obstetric ultrasound examinations are on average below the reference value. However, there is a significant number of cases in which the standards are exceeded, which should be a warning to the medical community.
Asunto(s)


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Ultrasonografía Prenatal / Obstetricia Tipo de estudio: Guía de Práctica Clínica Límite: Femenino / Humanos / Embarazo Idioma: Español Revista: Rev. chil. obstet. ginecol. (En línea) Asunto de la revista: Ginecologia / Obstetr¡cia Año: 2022 Tipo del documento: Artículo País de afiliación: Chile Institución/País de afiliación: Clínica Espacio EME/CL / Universidad Andrés Bello/CL / Universidad de Chile/CL / Universidad de Chile, Santiago/CL / Universidad de Valparaíso/CL

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Ultrasonografía Prenatal / Obstetricia Tipo de estudio: Guía de Práctica Clínica Límite: Femenino / Humanos / Embarazo Idioma: Español Revista: Rev. chil. obstet. ginecol. (En línea) Asunto de la revista: Ginecologia / Obstetr¡cia Año: 2022 Tipo del documento: Artículo País de afiliación: Chile Institución/País de afiliación: Clínica Espacio EME/CL / Universidad Andrés Bello/CL / Universidad de Chile/CL / Universidad de Chile, Santiago/CL / Universidad de Valparaíso/CL