Agreement analysis of stroke volume and cardiac output measurement between a oscillometric device and transthoracic echocardiogram in normotensive individuals: a preliminary report
Godoy, Alejandro; Contreras, Alejandro; Tabares, Aldo.
  • Godoy, Alejandro; Hospital Privado Universitario de Córdoba. Vascular Medicine and Thrombosis Service. Córdoba. AR
  • Contreras, Alejandro; Hospital Privado Universitario de Córdoba. Cardiology Service. Córdoba. AR
  • Tabares, Aldo; Hospital Privado Universitario de Córdoba. Vascular Medicine and Thrombosis Service. Córdoba. AR
Braz. J. Anesth. (Impr.) ; 73(4): 380-384, 2023. tab, graf
Artículo en Inglés | LILACS | ID: biblio-1447617
ABSTRACT
Abstract Introduction The evaluation of stroke volume (SV) is useful in research and patient care. To accomplish this, an ideal device should be noninvasive, continuous, reliable, and reproducible. The Mobil-O-Graph (MOG) is a noninvasive oscillometric matrix validated for measuring aortic and peripheral blood pressure, which through conversion algorithms can estimate hemodynamic parameters. Objectives To compare the MOG measurement of stroke volume, cardiac output, and cardiac index with the transthoracic echocardiogram (TTE). Methods Healthy volunteers aged 18 years or older were included. Two-dimensional TTEs were performed by a single operator. Subsequently, the measurement of noninvasive hemodynamics with MOG was performed with the operator blind to the results of the echocardiogram. Correlation analyses between stroke volume, cardiac output, and cardiac index parameters were performed. The degree of agreement between the methods was verified using the Bland-Altman method. Results A total of 38 volunteers were enrolled with a mean age of 27.6 ± 3.8 years; 21 (55%) were male The SV by TTE was 76.8 ± 19.5 mL and 75.7 ± 19.3 mL by MOG, Rho = 0.726, p< 0.0001. The CO by TTE was 5.04 ± 0.8 mL.min-1 and 5.1 ± 0.8 mL.min-1 by MOG Rho = 0.510, p= 0.001. Bland-Altman plots showed a good concordance between the two techniques. Conclusions Our study shows that the measurement of SV and CO by noninvasive hemodynamics with the MOG device offers a good concordance with the TTE with very few values beyond the confidence limits.
Asunto(s)


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Ecocardiografía / Hemodinámica Tipo de estudio: Ensayo Clínico Controlado Límite: Adulto / Femenino / Humanos / Masculino Idioma: Inglés Revista: Braz. J. Anesth. (Impr.) Año: 2023 Tipo del documento: Artículo País de afiliación: Argentina Institución/País de afiliación: Hospital Privado Universitario de Córdoba/AR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Ecocardiografía / Hemodinámica Tipo de estudio: Ensayo Clínico Controlado Límite: Adulto / Femenino / Humanos / Masculino Idioma: Inglés Revista: Braz. J. Anesth. (Impr.) Año: 2023 Tipo del documento: Artículo País de afiliación: Argentina Institución/País de afiliación: Hospital Privado Universitario de Córdoba/AR