Your browser doesn't support javascript.
loading
Inadequações dos Esfigmomanômetros Utilizados em Serviços de Urgência e Emergência de uma Grande Capital Brasileira / Inadequacies of Sphygmomanometers Used in Emergency Care Services in a Large Capital City in Brazil
Maia, Kleisson Antonio Pontes; Malachias, Marcus Vinícius Bolívar; Paiva, Isabela Viana de; Mariano, Rafael da Mota; Paiva, Rodrigo Viana de.
  • Maia, Kleisson Antonio Pontes; Faculdade de Ciências Médicas de Minas Gerais. Belo Horizonte - MG. BR
  • Malachias, Marcus Vinícius Bolívar; Faculdade de Ciências Médicas de Minas Gerais. Belo Horizonte - MG. BR
  • Paiva, Isabela Viana de; Faculdade de Ciências Médicas de Minas Gerais. Belo Horizonte - MG. BR
  • Mariano, Rafael da Mota; Faculdade de Ciências Médicas de Minas Gerais. Belo Horizonte - MG. BR
  • Paiva, Rodrigo Viana de; Faculdade de Ciências Médicas de Minas Gerais. Belo Horizonte - MG. BR
Int. j. cardiovasc. sci. (Impr.) ; 30(2): f:100-l:108, mar.-abr. 2017. tab
Artículo en Portugués | LILACS | ID: biblio-833892
RESUMO
Fundamento A hipertensão arterial é o principal fator de risco cardiovascular. A qualidade técnica dos esfigmomanômetros é condição fundamental para obtenção da correta medida da pressão arterial.

Objetivos:

Avaliar o perfil dos esfigmomanômetros disponíveis nos serviços de urgências de Belo Horizonte, Minas Gerais.

Métodos:

Realizamos um estudo transversal, observacional e não intervencionista para avaliar o perfil dos esfigmomanômetros disponíveis nos prontos atendimentos de adultos de hospitais públicos e privados do município de Belo Horizonte. Avaliamos 337 aparelhos de 25 hospitais, sendo 15 públicos (do total de 16) e 10 privados (do total de 12).

Resultados:

Foram observadas inadequações dos equipamentos em relação à validação pelo Instituto Nacional de Metrologia, Qualidade e Tecnologia ­ INMETRO, em 26% (88/337) dos equipamentos; calibração vencida em 39% (132/337) e não correspondência de marca aparelho/braçadeira em 54% (188/337). Em 13 dos 25 hospitais (52%), não havia disponibilidade de manguitos para braços de diferentes tamanhos além do padrão. Houve superioridade de adequação dos aparelhos aneroides e de mercúrio dos hospitais privados (p = 0,038 e p < 0,001, respectivamente) e dos eletrônicos nos públicos (p < 0,001).

Conclusão:

Conclui-se que 78% dos esfigmomanômetros disponíveis nos serviços de urgência/emergência de uma das maiores capitais brasileiras apresentavam inadequações técnicas, e em metade dos serviços não havia manguitos de diferentes tamanhos. Tal fato representa um alerta para a situação dos equipamentos disponíveis para atendimento da população no país
ABSTRACT

Background:

Hypertension is the main risk factor for cardiovascular diseases. Technical quality of sphygmomanometers is a prerequisite for the correct measurement of arterial pressure.

Objectives:

To evaluate sphygmomanometers available in emergency services in the city of Belo Horizonte, Brazil.

Methods:

We performed a cross-sectional, observational, non-interventional study to evaluate characteristics of the sphygmomanometers available in adult emergency services of public and private hospitals in the city of Belo Horizonte, Brazil. We evaluated 337 sphygmomanometers of 25 hospitals ­ 15 (of 16) public hospitals and 10 (of 12) private hospitals.

Results:

Twenty-six percent (88/337) of devices were considered inadequate regarding the INMETRO (National Institute of Metrology, Quality and Technology) standards, 39.2% (132/337) for calibration dates, and 54% (188/337) for the mismatching between cuff's and device's brands. In 13 of 25 hospitals (52%), there were no spare cuffs in different sizes for different arm circumferences. Higher adequacy was found for aneroid and mercury sphygmomanometers used in private hospitals (p = 0.038 and p < 0.001, respectively) and electronic devices used in public hospitals (p < 0.001) compared with others.

Conclusion:

Seventy-eight percent of sphygmomanometers available in emergency services had technical inadequacies, and half of these services had no spare cuffs in different sizes available. These findings serve as a warning of the conditions of the equipment used in healthcare services provided to the general population in Brazil
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Hospitales Privados / Esfigmomanometros / Hospitales Públicos / Hipertensión Tipo de estudio: Estudio de etiología / Estudio observacional / Estudio de prevalencia / Factores de riesgo País/Región como asunto: America del Sur / Brasil Idioma: Portugués Revista: Int. j. cardiovasc. sci. (Impr.) Asunto de la revista: Cardiología Año: 2017 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Faculdade de Ciências Médicas de Minas Gerais/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Hospitales Privados / Esfigmomanometros / Hospitales Públicos / Hipertensión Tipo de estudio: Estudio de etiología / Estudio observacional / Estudio de prevalencia / Factores de riesgo País/Región como asunto: America del Sur / Brasil Idioma: Portugués Revista: Int. j. cardiovasc. sci. (Impr.) Asunto de la revista: Cardiología Año: 2017 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Faculdade de Ciências Médicas de Minas Gerais/BR