Your browser doesn't support javascript.
loading
Desgaste dos instrumentos K3 e Protaper após simulação de uso clinico em canais curvos / Wear of K3 and ProTaper files after simulated clinical use in curved canals
Marceliano-Alves, Marília Fagury Videira; Santos, Mariana Diniz Bisi dos; Silva e Souza, Patrícia de Almeida Rodrigues.
  • Marceliano-Alves, Marília Fagury Videira; Universidade Federal do Pará. Faculdade de Odontologia. Belém. BR
  • Santos, Mariana Diniz Bisi dos; Universidade Federal do Pará. Faculdade de Odontologia. Belém. BR
  • Silva e Souza, Patrícia de Almeida Rodrigues; Centro Universitário do Pará. Belém. BR
RGO (Porto Alegre) ; 57(1): 13-18, jan.-mar. 2009. ilus
Artículo en Portugués | LILACS, BBO | ID: biblio-873760
RESUMO

Objetivo:

Avaliar o desgaste, após simulação de uso clínico, de instrumentos K3 (SybronEndo, Orange, CA, EUA) e ProTaper (Dentsply-Maillefer, Tulsa, Oklahoma, EUA) em 46 raízes mesiovestibulares de primeiros molares superiores humanos, recém-extraídos e com angulação média de 40º.

Métodos:

Os dentes foram divididos em a) K3 (SybronEndo, Orange, CA, EUA) (com bandas radiais) e b) ProTaper (Dentsply-Maillefer, Tulsa, Oklahoma, EUA) (sem bandas radiais). Os instrumentos foram avaliados em microscopia eletrônica de varredura antes do primeiro uso, para visualização de defeitos de fabricação, após 7 e 13 ciclos de instrumentação nos canais. A instrumentação foi realizada de acordo com as recomendações do fabricante, sendo K3 (SybronEndo, Orange, CA, EUA) e Protaper (Dentsply-Maillefer, Tulsa, Oklahoma, EUA). Os critérios para avaliação de desgaste foram defeitos (estiramento, encurtamento/ inversão das espiras, presença de crateras), trincas oblíquas e fratura. Após a avaliação em microscopia eletrônica de varredura antes do uso, foram constatadas em todos os instrumentos, farpas metálicas na crista das espiras e ranhuras oblíquas provenientes do processo de usinagem.

Resultados:

Após sete usos, no sistema K3 (SybronEndo, Orange, CA, EUA), todos os instrumentos apresentaram defeitos nos ângulos de corte e, 25/08 e 25/10 apresentaram trincas oblíquas no corpo da parte ativa levando à fratura dúctil desta última após o oitavo uso. No sistema ProTaper (Dentsply-Maillefer, Tulsa, Oklahoma, EUA), em sete usos, além de defeitos nos ângulos de corte, os instrumentos também apresentaram trincas, e a fratura rúptil ocorreu em F3 no 13º uso. Ao final do preparo foram observadas crateras ao longo da parte ativa dos instrumentos.

Conclusão:

Na análise final, foi evidenciado que os sistemas se equivalem quanto aos defeitos encontrados, independentemente do desenho da parte ativa, sendo proporcionais à quantidade de ciclos de uso, e que estes defeitos podem funcionar como...
ABSTRACT

Objective:

To assess wear, after simulated clinical use of K3 (SybronEndo, Orange, CA, EUA) and ProTaper files (Dentsply-Maillefer, Tulsa, Oklahoma, EUA) in 46 mesiovestibular roots of human maxillary first molars, recently extracted and with mean angulation of 40º.

Methods:

The teeth were divided as follows a) K3 (SybronEndo, Orange, CA, EUA) (with radial bands) and b) ProTaper (Dentsply-Maillefer, Tulsa, Oklahoma, EUA) (without radial bands). The files were assessed by scanning electronic microscopy before first use to visualize fabrication defects, and after 7 and 13 cycles of debridement in the canals. Debridement with files K3 (SybronEndo, Orange, CA, EUA) and ProTaper (Dentsply-Maillefer, Tulsa, Oklahoma, EUA) was performed in accordance with the manufacturers' recommendations. The criteria for wear assessment were defects (stretching, shortening/spiral inversion, presence of craters), oblique cracks and fracture. After assessment by SEM and before they were used, metal barbs on the crest of the spirals and oblique scratches due to the machining process were found in all the files.

Results:

In the K3 system (SybronEndo, Orange, CA, USA), after being used seven times, all the instruments showed cutting angle defects, and 25/08 and 25/10 had oblique cracks in the body of the active part leading to ductile fracture after being used for the eighth time. In the ProTaper system (Dentsply-Maillefer, Tulsa, Oklahoma, USA), after being used seven times, in addition to cutting angle defects, the files also presented cracks, and ruptile fracture occurred in F3 during use for the thirteenth time. At the end of preparation, craters were seen along the active part of the files.

Conclusion:

In the final analysis, it was seen that the system are equivalent with regard to the defects found, irrespective of the active part design, being proportional to the number of cycles of use, and these defects could act as points of greater fragility, favoring fracture.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Endodoncia Tipo de estudio: Estudios de evaluación / Guía de Práctica Clínica Límite: Humanos Idioma: Portugués Revista: RGO (Porto Alegre) Asunto de la revista: Odontología Año: 2009 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Centro Universitário do Pará/BR / Universidade Federal do Pará/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Endodoncia Tipo de estudio: Estudios de evaluación / Guía de Práctica Clínica Límite: Humanos Idioma: Portugués Revista: RGO (Porto Alegre) Asunto de la revista: Odontología Año: 2009 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Centro Universitário do Pará/BR / Universidade Federal do Pará/BR