Exploring cortical bone density through the ultrasound integrated reflection coefficient / Densidade óssea cortical analisada pelo coeficiente de reflexão ultrassônica
Acta ortop. bras
; 26(4): 255-259, July-Aug. 2018. tab, graf
Article
en En
| LILACS
| ID: biblio-973559
Biblioteca responsable:
BR1.1
ABSTRACT
ABSTRACT Objective:
This work evaluates the relationship between ultrasonic reflection and bone density from fourteen cylindrical bovine cortical bone samples (3.0-cm thick).Methods:
Twenty US reflection signals per sample were acquired along the bone surface (2.0-mm step). The Integrated Reflection Coefficient (IRC) from each signal was compared to Quantitative Computed Tomography (QCT).Results:
Seven IRC and QCT curves presented Pearson's Correlation R-values above 0.5. For weak correlation curves, QCT and IRC showed similar trends in several segments.Conclusion:
IRC was sensitive to bone density variation. Level of Evidence Experimental Study, Investigating a Diagnostic Test.RESUMO
RESUMO Objetivo:
Este estudo avalia a relação entre a reflexão ultrassônica e a densidade óssea de 14 amostras cilíndricas de osso cortical bovino (3,0 cm de espessura).Métodos:
Foi realizada a aquisição de 20 sinais de reflexão ultrassônica por amostra (passo de 2,0 mm), ao longo da superfície óssea. O Coeficiente de Reflexão Integrado (IRC) de cada sinal foi comparado por Tomografia Computadorizada Quantitativa (QCT).Resultados:
Sete curvas de IRC e QCT apresentaram valor de Correlação R de Pearson acima de 0,5. Para curvas de correlação fraca, QCT e IRC apresentaram tendências semelhantes em vários segmentos.Conclusão:
O IRC foi sensível à variação da densidade óssea. Nível de evidência Estudo Experimental, Investigação de Exame Diagnóstico.