Your browser doesn't support javascript.
loading
EDTA contamination in laboratory specimens-effect of an awareness campaign
JCPSP-Journal of the College of Physicians and Surgeons Pakistan. 2010; 20 (6): 405-407
en Inglés | IMEMR | ID: emr-98102
ABSTRACT
To measure the effects of awareness campaign on the frequency of EDTA contamination. A cohort study. The study was carried out in Combined Military Hospital, Peshawar, from October to December 2008. A cohort of 200 doctors, nurses and paramedical staff involved in sample collection for laboratory investigations were selected for this three phased study. In all samples with plasma potassium level>6.0 mmol/L, plasma Ca++ was measured for one month before, during and after a campaign of correct filling order of sample tubes for two weeks. Renal function test and LDH were also measured to rule out chronic renal failure and in-vitro haemolysis, respectively. Fisher's exact test was used for comparison of frequency of hyperkalemia. There was a significant decrease [75%, p<0.01] in the cases of spurious hyperkalaemia and hypocalcaemia. EDTA contamination is a common and important source of pre-analytical error which can be prevented to some extent by education of medical and nursing staff
Asunto(s)
Buscar en Google
Índice: IMEMR (Mediterraneo Oriental) Asunto principal: Laboratorios de Hospital / Recolección de Muestras de Sangre / Educación en Salud / Ácido Edético / Cuerpo Médico de Hospitales / Anticoagulantes Límite: Humanos Idioma: Inglés Revista: J. Coll. Physicians Surg. Pak. Año: 2010

Similares

MEDLINE

...
LILACS

LIS

Buscar en Google
Índice: IMEMR (Mediterraneo Oriental) Asunto principal: Laboratorios de Hospital / Recolección de Muestras de Sangre / Educación en Salud / Ácido Edético / Cuerpo Médico de Hospitales / Anticoagulantes Límite: Humanos Idioma: Inglés Revista: J. Coll. Physicians Surg. Pak. Año: 2010