Avaliação do aparelho automático oscilométrico de medida de pressão arterial DIXTAL DX-2710 pelos protocolos de validação da British Hypertension Society (BHS) e Association for de Advancement of Medical Instrumentation (AAMI) / Assessment of the DIXTAL DX-2710 automated oscillometric device for blood pressure measurement with the validation protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI)
Mano, Gisele M. P; Souza, Viviane F; Pierin, Angela M. G; Lima, Josiane C; Ignes, Edna C; Ortega, Kátia C; Mion Júnior, Décio.
  • Mano, Gisele M. P; Universidade de Säo Paulo. São Paulo. BR
  • Souza, Viviane F; Universidade de Säo Paulo. São Paulo. BR
  • Pierin, Angela M. G; Universidade de Säo Paulo. São Paulo. BR
  • Lima, Josiane C; Universidade de Säo Paulo. São Paulo. BR
  • Ignes, Edna C; Universidade de Säo Paulo. São Paulo. BR
  • Ortega, Kátia C; Universidade de Säo Paulo. São Paulo. BR
  • Mion Júnior, Décio; Universidade de Säo Paulo. São Paulo. BR
Arq. bras. cardiol ; 79(6): 601-610, dez. 2002. tab, graf
Artículo en Portugués, Inglés | LILACS | ID: lil-326147
ABSTRACT

OBJECTIVE:

To assess the Dixtal DX2710 automated oscillometric device used for blood pressure measurement according to the protocols of the BHS and the AAMI.

METHODS:

Three blood pressure measurements were taken in 94 patients (53 females 15 to 80 years). The measurements were taken randomly by 2 observers trained to measure blood pressure with a mercury column device connected with an automated device. The device was classified according to the protocols of the BHS and AAMI.

RESULT:

The mean of blood pressure levels obtained by the observers was 148±38/93±25 mmHg and that obtained with the device was 148±37/89±26 mmHg. Considering the differences between the measurements obtained by the observer and those obtained with the automated device according to the criteria of the BHS, the following classification was adopted "A" for systolic pressure (69 percent of the differences < 5; 90 percent < 10; and 97 percent < 15 mmHg); and "B" for diastolic pressure (63 percent of the differences < 5; 83 percent < 10; and 93 percent < 15 mmHg). The mean and standard deviation of the differences were 0±6.27 mmHg for systolic pressure and 3.82±6.21 mmHg for diastolic pressure.

CONCLUSION:

The Dixtal DX2710 device was approved according to the international

recommendations:

Asunto(s)
Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Presión Sanguínea / Monitoreo Ambulatorio / Hipertensión Tipo de estudio: Ensayo Clínico Controlado / Estudios de evaluación / Factores de riesgo Límite: Adolescente / Adulto / Aged80 / Femenino / Humanos / Masculino Idioma: Inglés / Portugués Revista: Arq. bras. cardiol Asunto de la revista: Cardiología Año: 2002 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade de Säo Paulo/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Presión Sanguínea / Monitoreo Ambulatorio / Hipertensión Tipo de estudio: Ensayo Clínico Controlado / Estudios de evaluación / Factores de riesgo Límite: Adolescente / Adulto / Aged80 / Femenino / Humanos / Masculino Idioma: Inglés / Portugués Revista: Arq. bras. cardiol Asunto de la revista: Cardiología Año: 2002 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade de Säo Paulo/BR