Intercomparação e calibração de medidores de atividade utilizados em serviços de medicina nuclear / Intercomparison and calibration of dose calibrators used in nuclear medicine facilities
Costa, Alessandro Martins da; Caldas, Linda V. E.
  • Costa, Alessandro Martins da; Comissão Nacional de Energia Nuclear. Instituto de Pesquisas Energéticas e Nucleares. São Paulo. BR
  • Caldas, Linda V. E; s.af
Radiol. bras ; 36(5): 293-297, set.-out. 2003. tab, graf
Artículo en Portugués | LILACS | ID: lil-351426
RESUMO

OBJETIVO:

O objetivo deste trabalho foi estabelecer um padrão de trabalho para intercomparação e calibração de medidores de atividade (calibradores de dose) utilizados na maioria dos serviços de medicina nuclear, para determinação da atividade dos radionuclídeos administrados aos pacientes nos exames específicos ou nos procedimentos terapêuticos.

MATERIAIS E MÉTODOS:

Foi utilizado um calibrador de dose comercial, um conjunto de fontes radioativas padrões, além de seringas, frascos e ampolas contendo soluções de radionuclídeos utilizados em medicina nuclear.

RESULTADOS:

O calibrador de dose comercial foi calibrado com soluções de radionuclídeos utilizados em medicina nuclear. Os testes simples do instrumento, tais como o teste de linearidade e da variação da resposta com o volume da fonte a uma concentração de atividade constante, foram descritos e realizados.

CONCLUSÃO:

O instrumento estudado pode agora ser utilizado como sistema de referência para intercomparação e calibração de outros instrumentos medidores de atividade, como um método de controle da qualidade de calibradores de dose utilizados em serviços de medicina nuclear.
ABSTRACT

OBJECTIVE:

The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. MATERIALS AND

METHODS:

A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work.

RESULTS:

The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed.

CONCLUSION:

This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities
Asunto(s)
Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Radioisótopos / Radiometría / Calibración / Medicina Nuclear Tipo de estudio: Estudio pronóstico Idioma: Portugués Revista: Radiol. bras Asunto de la revista: Diagnóstico por Imagen / Radiología Año: 2003 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Comissão Nacional de Energia Nuclear/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Radioisótopos / Radiometría / Calibración / Medicina Nuclear Tipo de estudio: Estudio pronóstico Idioma: Portugués Revista: Radiol. bras Asunto de la revista: Diagnóstico por Imagen / Radiología Año: 2003 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Comissão Nacional de Energia Nuclear/BR