Determinación y análisis comparativo de gasto cardíaco en enfermos críticos mediante Doppler transesofágico y termodilución en bolos: Experiencia preliminar / Comparison of cardiac output measurement in critical patients using transesophageal Doppler or thermodilution: A preliminary report
Andresen H., Max; Henríquez V., Mauricio; Mercado F., Marcelo; Farías G., Gonzalo; Castillo S., Carmen; Benítez G., Carlos; Dougnac L., Alberto.
  • Andresen H., Max; Pontificia Universidad Católica de Chile. Facultad de Medicina. Hospital Clínico. Unidad de Tratamiento Intensivo Médico. Santiago. CL
  • Henríquez V., Mauricio; Pontificia Universidad Católica de Chile. Facultad de Medicina. Hospital Clínico. Santiago. CL
  • Mercado F., Marcelo; Pontificia Universidad Católica de Chile. Facultad de Medicina. Hospital Clínico. Unidad de Tratamiento Intensivo Médico. Santiago. CL
  • Farías G., Gonzalo; Pontificia Universidad Católica de Chile. Facultad de Medicina. Hospital Clínico. Santiago. CL
  • Castillo S., Carmen; Pontificia Universidad Católica de Chile. Facultad de Medicina. Hospital Clínico. Santiago. CL
  • Benítez G., Carlos; Pontificia Universidad Católica de Chile. Facultad de Medicina. Hospital Clínico. Santiago. CL
  • Dougnac L., Alberto; Pontificia Universidad Católica de Chile. Facultad de Medicina. Hospital Clínico. Unidad de Tratamiento Intensivo Médico. Santiago. CL
Rev. méd. Chile ; 133(7): 761-766, jul. 2005. ilus, tab
Artículo en Español | LILACS | ID: lil-429134
ABSTRACT
Background: Cardiac output can be measured non invasively by transesophageal Doppler. This is an alternative to measure it by thermodilution with a catheter in the pulmonary artery. Aim: To compare both methods of cardiac output measurement. Material and methods: Simultaneous measurement of cardiac output by transesophageal Doppler and thermodilution with a catheter in the pulmonary artery in four male critical patients, aged 60±12 years, hospitalized in a University Hospital. The Bland and Altman method to compare the concordance between two measurements, was used. Results: Forty measurements were performed. The results of both methods had a correlation coefficient of 0.98. According to the Bland and Altman method, the difference between both methods was -0.5 L with a precision of 0.52 L/min (95% confidence interval -1.51 to 0.52 L/min). Considering that a change between two sequential measurements is considered significant when the difference is more than 15%, both measurements agreed in 83% of cases, that there was a change in cardiac output. Conclusions:Transesophageal Doppler is a promising non invasive technique to measure cardiac output in critical care patients. It becomes a valid alternative to the thermodilution technique. This preliminary experience must be confirmed in a larger series.
Asunto(s)
Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Termodilución / Cateterismo de Swan-Ganz / Gasto Cardíaco / Ecocardiografía Doppler / Ecocardiografía Transesofágica Tipo de estudio: Evaluación Económica en Salud / Estudio observacional Límite: Humanos / Masculino Idioma: Español Revista: Rev. méd. Chile Asunto de la revista: Medicina Año: 2005 Tipo del documento: Artículo País de afiliación: Chile Institución/País de afiliación: Pontificia Universidad Católica de Chile/CL

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Termodilución / Cateterismo de Swan-Ganz / Gasto Cardíaco / Ecocardiografía Doppler / Ecocardiografía Transesofágica Tipo de estudio: Evaluación Económica en Salud / Estudio observacional Límite: Humanos / Masculino Idioma: Español Revista: Rev. méd. Chile Asunto de la revista: Medicina Año: 2005 Tipo del documento: Artículo País de afiliación: Chile Institución/País de afiliación: Pontificia Universidad Católica de Chile/CL